Acid Digestion Fume Hood

Send Inquiry
Acid Digestion Fume Hood
Details
Perchloric Acid digestion fume hood overcomes the shortcomings of traditional fume hoods, which are prone to rust, yellowing, and cracking in high-temperature, concentrated acid environments, and offers excellent acid and alkali resistance. It is suitable for pharmaceutical lab, food lab, and semiconductor lab.
Category
Ducted Fume Hoods
Share to
Description

Base cabinet valve function selection

 

 

1. water inlet valve of pump When the water inlet valve of the Acid digestion fume hood's pump is in use, it is normally open and must not be closed, except when replacing the pump.

2. water inlet valve of the water tank When the water inlet valve of the water tank is in use, the valve is normally open and not allowed to close (except for the need to close when replacing the float ball).

3. blow-down valve When the blow down valve is in use, it is normally closed and not allowed to be opened (except for the need to replace the absorbed liquid in the water tank and open when blow down).

4. small sink valve on the table Under the use of small sink valve on the table, the valve is normally closed and not allowed to be opened (except when cleaning the hanging cabinet of fume hood separately).

Ducted Fume Hood

5. spraying valve of the upper cabinet When the spray valve in the upper section of the Acid digestion fume hood is in use, it should normally remain open and must not be closed, except when the water tank is being drained.

 

SPECIFICATIONS

 

 

Item

Details

Dimensions

1200/1500/1800mm (L) × 1100mm (D) × 2300mm (H)

Main Material

Porcelain white PP, 8mm thick, resistant to acid and alkali corrosion

Worktop

PP material

Power Supply

AC 220V ± 10%, 50/60Hz

Integrated Features

Built-in flushing system

 

TECHNICAL DATA

MODEL

HM-PFH-120

HM-PFH-150

HM-PFH-180

Outer Dimension
(LxWxH)

1200*1050*2350

1500*1050*2350

1800*1050*2350

Working Area Dimension(LxWxH)

1000*600*1100mm

1300*600*1100mm

1600*600*1100mm

Max. door opening

700mm(Standard vertical adjustment door)

Worktop

Ceramic;PTFE

Sash

5mm transparent tempered glass

Air hood

250mm diameter air hood

 

SUMMARIZEUSAGE PATTERN

 

 

  • normal spray condition: Open valve 1/2/5 and close valve 3/4
  • sewage discharge condition: Open valve 1/3 and close valve 2/4/5
  •  Hanging cabinet cleaning (manual cleaning, no need to start the water pump) Open valve 1/2, 4/5, close valve 3

 

The Water Tank of Base Cabinet

 
  1. Reagent inlet: manually remove the white top cover, and pour the proportionally prepared absorption reagent into the water tank. Just put the lid back on.
  2. Sampling tap: the sampling tap manually turns the handle of the tap, and immediately turn off after collecting the liquid sample

Acid Digestion

Water Level of Water tank

 

 

Water Level of Water tank

  1. work water level: Manually adjust the liquid level of the water tank to the working liquid level by adjusting the adjustment screw on the float ball, when the equipment power off.
  2. Max water level: The circulating liquid will return to the water tank, and the liquid level will not exceed this level.
  3. The lowest permissible water level is the minimum limit. Under no circumstances is the circulating pump of the Acid digestion fume hood system to be started below this level. An excessively low level will cause the impeller to run dry, resulting in pump damage. This includes during drainage: closely monitor the tank level, and if it falls below this point, power must be cut off immediately.